DOT breath alcohol testing follows the procedures in 49 CFR Part 40, Subparts J through N. Only a qualified screening test technician (STT) or breath alcohol technician (BAT) may conduct a DOT alcohol test. An STT can conduct only screening tests. A BAT can conduct both screening and confirmation tests (49 CFR 40.211).
Approved Devices
A screening test may be done on an alcohol screening device (ASD) or an evidential breath testing device (EBT). Both must appear on ODAPC's lists of approved devices. An ASD, such as a saliva device, can be used only for screening tests and never for confirmation tests (49 CFR 40.229).
Only an approved EBT may be used for a confirmation test. The EBT must print each result in triplicate (or three consecutive identical copies), assign a unique number to each test, test an air blank and perform an external calibration check (49 CFR 40.231).
Qualification training must cover the particular ASD or EBT the technician will use. After training, a BAT must complete seven consecutive error-free mock tests, and an STT must complete five (49 CFR 40.213).
The Testing Site
The testing site must give the employee visual and aural privacy so that unauthorized persons cannot see or hear test results (49 CFR 40.221). The technician may test only one employee at a time. When a BAT will do both tests, the full screening and confirmation process for one employee must be finished before the next employee's screening begins (49 CFR 40.223).
The Screening Test
The technician checks the employee's photo ID, explains the procedure and completes Step 1 of the DOT Alcohol Testing Form (ATF). The employee then completes and signs Step 2. Refusing to sign Step 2 is a refusal to test. If the employee also has a DOT drug test, the alcohol test should be completed first whenever practicable (49 CFR 40.241).
On a breath device, the technician opens an individually wrapped or sealed mouthpiece in view of the employee. The employee blows steadily and forcefully for at least six seconds or until the device shows that enough breath was collected (49 CFR 40.243). On a saliva ASD, the technician shows the employee the expiration date, opens the sealed package in front of the employee and reads the result no sooner than the manufacturer allows and always within 15 minutes of the test (49 CFR 40.245).
A screening result below 0.02 ends the test. The technician signs the ATF and sends the result to the Designated Employer Representative (DER). A result of 0.02 or higher requires a confirmation test (49 CFR 40.247).
The 15-Minute Waiting Period
The confirmation test cannot start until at least 15 minutes after the screening test is completed. It should begin as soon as possible after that, and no more than 30 minutes after the screening test. The employee must be observed during the wait (49 CFR 40.251).
The BAT tells the employee not to eat, drink, put anything in the mouth or belch during the wait. The wait prevents mouth alcohol from causing an artificially high reading. The confirmation test goes ahead at the end of the wait even if the employee did not follow these instructions (49 CFR 40.251(a)(2)).
A confirmation test done before the 15 minutes are up is a fatal flaw, and the test must be cancelled (49 CFR 40.267). Starting after 30 minutes does not invalidate the test, but it may be a regulatory violation (49 CFR 40.251(g)).
The Confirmation Test
Before the confirmation test, the BAT runs an air blank on the EBT in front of the employee. The reading must be 0.00. If it is not, the BAT runs a second air blank. If the second reading is also above 0.00, the EBT must be taken out of service (49 CFR 40.253).
The BAT then opens a new individually wrapped or sealed mouthpiece in view of the employee. The BAT and the employee both read the unique test number on the EBT. The employee blows for at least six seconds, and the BAT shows the employee the displayed result and the printed result (49 CFR 40.253).
If the confirmation result is 0.02 or higher, the employee signs Step 4 of the ATF, and the BAT must notify the DER immediately and directly, not through a C/TPA. If it is below 0.02, nothing further is required of the employee (49 CFR 40.255).
Refusals
Refusals include failing to provide enough breath or saliva, failing to sign Step 2 of the ATF and failing to cooperate with any part of the testing process. The technician notes the conduct on the ATF and notifies the DER right away. The technician does not decide whether a refusal occurred. That decision belongs to the employer alone (49 CFR 40.261 and 40.355(i)).
Hygiene and Device Care
Several Part 40 steps also support hygiene. Mouthpieces must be individually wrapped or sealed and opened in view of the employee (49 CFR 40.243 and 40.253). A technician who places a saliva device in an employee's mouth must wear single-use gloves and change them after each test. Saliva devices, swabs and gloves are never reused (49 CFR 40.245).
EBT users must follow the manufacturer's instructions for use and care, including external calibration checks at the required intervals (49 CFR 40.233). Any cleaning product used on a device should be one the manufacturer's instructions allow.
The full rules are in 49 CFR Part 40 on eCFR.